Our Commitment

Quality You
Can Rely On

At Adensys, we go beyond delivering dental parts. We deliver peace of mind.

Traceability That Builds Trust

Every Adensys rotor is marked with a unique serial number, ensuring full traceability from production to delivery. This allows us to guarantee quality, simplify warranty management, and give you confidence that if an issue ever arises, we can track and resolve it quickly.

Technical Properties

Although we fulfil all of the ISO-stipulated requirements listed above, we go above and beyond whenever possible. Specifically, this means that our cartridge products have technical properties that match or exceed the properties you would find from the original equipment manufacturer.

High Retention Force – ensures safety for patients and long spindle lifetime

Force required to extract the bur from the chuck

>30N

Low Thumb Force – makes for easy operation

Force required to open the chuck mechanism using the push button

25N – 30N

Low Bur Runout – increases bearing lifetime and improves precision operation of handpiec

The radial deviation of a bur, from completely concentric, when turned measured
6cm from the end of the rotor‘s axis part

< 0.015mm

High Part Hardness – increase bearing and chuck lifetime

The hardness of all parts, including axis on which the bearings are mounted

60-65 HRC

Low Imbalance – reduces handpiece noise and increases rotor lifetime

The amount of rotor imbalance, expressed as a mass at a certain distance
from the centre of rotation

< 0.2mgmm

Traceability Through Serial Number

All rotor products are delivered with a unique serial number (at the front of the spindle), which allows for smoother management of guarantee claims and allows for full traceability if any quality issues arise in the market. Measured technical properties for each rotor are recorded and archived by ADENSYS.

ISO Compliance

In order to fulfil the vision spelled out in our quality philosophy, we have to ensure that we deliver the top technical characteristics for all ADENSYS products. In the case of our rotor assemblies, this means that absolutely all requirements of globally recognised standards are respected. Our rotors fulfil the requirements for the following ISO norms:

ISO 7785-1 & 7785-2

Technical Requirements for air- and motor-driven dental handpieces

ISO 1797

Specifications for the burs and mandrels that are used with dental handpieces

ISO 9687

Description of the graphical symbols to be used for dental equipment

ISO 13060

Process description for sterilisation of medical
devices

ISO 14457

The norm for dental hand pieces and motors, which replaced ISO 7785 and ISO 13294

This site is registered on wpml.org as a development site. Switch to a production site key to remove this banner.